After analysis, this is not a bug but simply a sorting according to Anatomical Direction that is forced in MPR mode.
Anatomical Direction is defined in DICOM the x-axis is increasing to the left hand side of the patient. The y-axis is increasing to the posterior side of the patient. The z-axis is increasing toward the head of the patient.
Describe the bug. What happened?
Hi,
I am using weasis 4.4.0 with Windows 11.
Here's what I do:
The slice number of a sequence in MPR mode seems reversed:
I am not sure if this is a bug or by design.
What version of Weasis are you running?
4.4.0
On which system the problem occurs?
Windows
Relevant log output
No response
Additional contextual elements
No response